Complete Prevention of Bubbles in a PDMS-Based Digital PCR Chip with a Multifunction Cavity
Shiyuan Gao,
Tiegang Xu,
Lei Wu,
Xiaoyue Zhu,
Xuefeng Wang,
Ying Chen,
Gang Li,
Xinxin Li
Affiliations
Shiyuan Gao
State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China
Tiegang Xu
State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China
Lei Wu
State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China
Xiaoyue Zhu
Metabolomics Center, Haixia Institute of Science and Technology, School of Future Technology, Fujian Agriculture and Forestry University, Fuzhou 350002, China
Xuefeng Wang
State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China
Ying Chen
State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China
Gang Li
Key Laboratory of Optoelectronic Technology and Systems, Ministry of Education, Defense Key Disciplines Lab of Novel Micro-Nano Devices and System Technology, Chongqing University, Chongqing 400044, China
Xinxin Li
State Key Laboratory of Transducer Technology, Shanghai Institute of Microsystem and Information Technology, Chinese Academy of Sciences, Shanghai 200050, China
In a chamber-based digital PCR (dPCR) chip fabricated with polydimethylsiloxane (PDMS), bubble generation in the chambers at high temperatures is a critical issue. Here, we found that the main reason for bubble formation in PDMS chips is the too-high saturated vapor pressure of water at an elevated temperature. The bubbles should be completely prevented by reducing the initial pressure of the system to under 13.6 kPa to eliminate the effects of increased-pressure water vapor. Then, a cavity was designed and fabricated above the PCR reaction layer, and Parylene C was used as a shell covering the chip. The cavity was used for the negative generator in sample loading, PDMS degassing, PCR solution degassing in the digitization process and water storage in the thermal reaction process. The analysis was confirmed and finally achieved a desirable bubble-free, fast-digitization, valve-free and no-tubing connection dPCR.