Analysis of Strain Distribution in Common Clinical Designs of Posterior Implant-Supported Fixed Partial Restorations: Comparison between Six Configurations
Abstract
The configuration of implant-supported prostheses is considered to influence the magnitude of stress concentrations, affecting their survival rate. The purpose of this study is to determine, through strain gauge measurements during load application, the dispersion and magnitude of strain concentrations in different implant-supported prosthesis designs. All designs matched those commonly used in posterior partially edentulous states. Three implants were inserted into an epoxy resin model (PLM-4B Vishay Measurements Group Inc., Raleigh, NC, USA), allowing for the delivery of three- and four-unit crowns in different cemented configurations. Loads were applied at vertical and oblique directions over the cast crowns in six different configurations representing various posterior partially edentulous restorations. The readings from the strain gauges adhered to the implant necks’ presented data on implant strain. Prostheses including cantilevers showed the highest strain among the three-unit prostheses within the prosthetic complex, and three single units showed the least (8133 µs vs. 201 µs, respectively). Angulated load application also had a role in amplifying the strains recorded, resulting in total strains of between 3.5 and 20 times higher than during vertical loading in all configurations. It can be concluded that the configuration of implant-fixed partial prosthesis changes the loads engaging the restoration, the implant, and, probably, the supporting bone.