Physics and Imaging in Radiation Oncology (Jan 2024)

A look-up-table development to facilitate CT simulation of MR-Linac treatment

  • Neil Hughes,
  • Travis C. Salzillo,
  • Sastry Vedam,
  • Tze Yee Lim,
  • Xin Wang,
  • He Wang,
  • Mustefa Mohammedsaid,
  • Clifton D. Fuller,
  • Jihong Wang,
  • Jinzhong Yang

Journal volume & issue
Vol. 29
p. 100524

Abstract

Read online

While current MR-Linac (MRL) treatment workflows utilize a large table overlay during CT simulation to convert indexing between the two machines, we developed a look-up-table (LUT) as an alternative approach. After populating the LUT, index conversion factors were verified at three separate table locations. The resultant root-mean-square isocenter shifts on the MRL were 0.04/0.08 cm, 0.08/0.07 cm, and 0.09/0.08 cm with/without using the table overlay during simulation in the lateral, longitudinal, and vertical directions, respectively, which is within registration tolerance. Clinical implementation of the LUT has resulted in a more efficient MRL treatment workflow while maintaining accurate patient setup.