PLoS ONE (Jan 2020)

Analysis and modeling of coolants and coolers for specimen transportation.

  • David E Lowe,
  • Gerald Pellegrini,
  • Elizabeth LeMasters,
  • Andrew J Carter,
  • Zachary P Weiner

DOI
https://doi.org/10.1371/journal.pone.0231093
Journal volume & issue
Vol. 15, no. 4
p. e0231093

Abstract

Read online

Maintaining cold chain while transporting medical supplies and samples is difficult in remote settings. Failure to maintain temperature requirements can lead to degraded sample quality and inaccuracies in sample analysis. We performed a systematic analysis on different types of transport coolers (polystyrene foam, injection-molded, and rotational molded) and transport coolants (ice, cold packs, frozen water bottles) frequently in use in many countries. Polystyrene foam coolers stayed below our temperature threshold (6°C) longer than almost all other types of coolers, but were not durable. Injection-molded coolers were durable, but warmed to 6°C the quickest. Rotational molded coolers were able to keep temperatures below our threshold for 24 hours longer than injection molded coolers and were highly durable. Coolant systems were evaluated in terms of cost and their ability to maintain cold temperatures. Long lasting commercial cold packs were found to be less cost effective and were below freezing for the majority of the testing period. Frozen plastic water bottles were found to be a reusable and economical choice for coolant and were only below freezing briefly. Finally, we modeled the coolers performance at maintaining internal temperatures below 6°C and built a highly accurate linear model to predict how long a cooler will remain below 6°C. We believe this data may be useful in the planning and design of specimen transportation systems in the field, particularly in remote or resource limited settings.