Sensors (Aug 2019)

Noninvasive Assessment of Aortic Pulse Wave Velocity by the Brachial Occlusion-Cuff Technique: Comparative Study

  • Vratislav Fabian,
  • Lukas Matera,
  • Kristyna Bayerova,
  • Jan Havlik,
  • Vaclav Kremen,
  • Jan Pudil,
  • Pavol Sajgalik,
  • David Zemanek

DOI
https://doi.org/10.3390/s19163467
Journal volume & issue
Vol. 19, no. 16
p. 3467

Abstract

Read online

Cardiovascular diseases are one of most frequent cause of morbidity and mortality in the world. There is an emerging need for integrated, non-invasive, and easy-to-use clinical tools to assess accurately cardiovascular system primarily in the preventative medicine. We present a novel design for a non-invasive pulse wave velocity (PWV) assessment method integrated in a single brachial blood pressure monitor allowing for up to 100 times more sensitive recording of the pressure pulsations based on a brachial occlusion-cuff (suprasystolic) principle. The monitor prototype with built-in proprietary method was validated with a gold standard reference technique SphygmoCor VX device. The blood pressure and PWV were assessed on twenty-five healthy individuals (9 women, age (37 ± 13) years) in a supine position at rest by a brachial cuff blood pressure monitor prototype, and immediately re-tested using a gold standard method. PWV using our BP monitor was (6.67 ± 0.96) m/s compared to PWV determined by SphygmoCor VX (6.15 ± 1.01) m/s. The correlation between methods using a Pearson’s correlation coefficient was r = 0.88 (p < 0.001). The study demonstrates the feasibility of using a single brachial cuff build-in technique for the assessment of the arterial stiffness from a single ambulatory blood pressure assessment.