Respiratory Research (Oct 2024)

Induction of subject-ventilator asynchrony by variation of respiratory parameters in a lung injury model in pigs

  • Xi Ran,
  • Martin Scharffenberg,
  • Jakob Wittenstein,
  • Mark Leidermann,
  • Andreas Güldner,
  • Thea Koch,
  • Marcelo Gama de Abreu,
  • Robert Huhle

DOI
https://doi.org/10.1186/s12931-024-02984-y
Journal volume & issue
Vol. 25, no. 1
pp. 1 – 14

Abstract

Read online

Abstract Background Subject-ventilator asynchrony (SVA) was shown to be associated with negative clinical outcomes. To elucidate pathophysiology pathways and effects of SVA on lung tissue histology a reproducible animal model of artificially induced asynchrony was developed and evaluated. Methods Alterations in ventilator parameters were used to induce the three main types of asynchrony: ineffective efforts (IE), auto-triggering (AT), and double-triggering (DT). Airway flow and pressure, as well as oesophageal pressure waveforms, were recorded, asynchrony cycles were manually classified and the asynchrony index (AIX) was calculated. Bench tests were conducted on an active lung simulator with ventilator settings altered cycle by cycle. The developed algorithm was evaluated in three pilot experiments and a study in pigs ventilated for twelve hours with AIX = 25%. Results IE and AT were induced reliably and fail-safe by end-expiratory hold and adjustment of respiratory rate, respectively. DT was provoked using airway pressure ramp prolongation, however not controlled specifically in the pilots. In the subsequent study, an AIX = 28.8% [24.0%-34.4%] was induced and maintained over twelve hours. Conclusions The method allows to reproducibly induce and maintain three clinically relevant types of SVA observed in ventilated patients and may thus serve as a useful tool for future investigations on cellular and inflammatory effects of asynchrony.

Keywords