Materials (Mar 2022)

Discolouring 3D Gel Dosimeter for UV Dose Distribution Measurements

  • Malwina Jaszczak,
  • Elżbieta Sąsiadek-Andrzejczak,
  • Marek Kozicki

DOI
https://doi.org/10.3390/ma15072546
Journal volume & issue
Vol. 15, no. 7
p. 2546

Abstract

Read online

This work reports on a new TBO–Pluronic F–127 three-dimensional (3D) gel dosimeter for UV light dose distribution measurements. The optimal gel composition was found to be 60 µM Toluidine Blue O (TBO), which acts as a UV-sensitive compound; 5% w/w hydrogen peroxide (H2O2), which is necessary for initiation of TBO photodegradation and 25% w/w poly(ethylene oxide)-block-poly(propylene oxide)-block-poly(ethylene oxide) (Pluronic F–127), which forms a physical gel matrix. The dosimeter becomes discoloured when exposed to UV radiation and a discolouration is the more intense, the higher the absorbed dose is. The samples after irradiation with UVA, UVB and UVC radiation were measured using UV-Vis spectrophotometry to obtain the basic dose–response characteristic of the dosimeter, including dose sensitivity, linear and dynamic dose range, threshold dose, stability over time and dose–response for fractioned and non-fractioned doses. Additionally, the TBO–Pluronic F–127 gel dosimeter was investigated for spatial stability and the ability to measure the dose distribution of UV radiation. The results obtained indicate that the TBO–Pluronic F–127 dosimeter is a promising UV sensor and 2D/3D UV dosimeter.

Keywords