BMC Medical Imaging (Aug 2021)

Micro-CT yields high image quality in human fetal post-mortem imaging despite maceration

  • Ian Craig Simcock,
  • Susan Cheng Shelmerdine,
  • Dean Langan,
  • Guy Anna,
  • Neil James Sebire,
  • Owen John Arthurs

DOI
https://doi.org/10.1186/s12880-021-00658-5
Journal volume & issue
Vol. 21, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Background Current clinical post-mortem imaging techniques do not provide sufficiently high-resolution imaging for smaller fetuses after pregnancy loss. Post-mortem micro-CT is a non-invasive technique that can deliver high diagnostic accuracy for these smaller fetuses. The purpose of the study is to identify the main predictors of image quality for human fetal post-mortem micro-CT imaging. Methods Human fetuses were imaged using micro-CT following potassium tri-iodide tissue preparation, and axial head and chest views were assessed for image quality on a Likert scale by two blinded radiologists. Simple and multivariable linear regression models were performed with demographic details, iodination, tissue maceration score and imaging parameters as predictor variables. Results 258 fetuses were assessed, with median weight 41.7 g (2.6–350 g) and mean gestational age 16 weeks (11–24 weeks). A high image quality score (> 6.5) was achieved in 95% of micro-CT studies, higher for the head (median = 9) than chest (median = 8.5) imaging. The strongest negative predictors of image quality were increasing maceration and body weight (p < 0.001), with number of projections being the best positive imaging predictor. Conclusions High micro-CT image quality score is achievable following early pregnancy loss despite fetal maceration, particularly in smaller fetuses where conventional autopsy may be particularly challenging. These findings will help establish clinical micro-CT imaging services, addressing the need for less invasive fetal autopsy methods.

Keywords