Indian Journal of Ophthalmology (Jan 2000)

Visual field assessment in glaucoma: Comparative evaluation of manual Kinetic Goldmann Perimetry and Automated Static Perimetry

  • Agarwal Harish,
  • Gulati V.,
  • Sihota Ramanjit

Journal volume & issue
Vol. 48, no. 4
pp. 301 – 6

Abstract

Read online

Purpose: To compare the detection and assessment of progression of visual field defects in primary open-angle glaucoma with manual suprathreshold perimetry on Goldmann perimeter and automated static threshold perimetry on Humphery visual field (HVF) analyzer. Methods: 105 eyes of 54 patients of primary open-angle glaucoma were followed up with 3-monthly perimetry on Goldmann perimeter and HVF analyzer, for a period of 9 months. Results: HVF analyzer picked up visual field defects in 48 (46%) eyes whereas Goldmann perimeter picked up visual field defects in 26 (25%) eyes. HVF analyzer demonstrated progression in 14 eyes whereas Goldmann perimeter detected progression in 7 eyes during follow up of 9 months. Conclusions: HVF analyzer is superior to Goldmann perimeter to document and to demonstrate progression of visual field defects in primary open-angle glaucoma.