Using Real-Time PCR Fluorescence Reaction Values to Improve SARS-CoV-2 Virus Detection and Benefit Clinical Decision-Making

  • Wan-Wen Yang,
  • Chin-Wen Hsu,
  • Yu-Ju Chan,
  • Shih-Bin Su,
  • I-Jung Feng,
  • Chia-Yi Hou,
  • Chien-Yuan Huang

DOI
https://doi.org/10.3390/life13030683
Journal volume & issue
Vol. 13, no. 3
p. 683

Abstract

Read online

This study aimed to compare the SARS-CoV-2 nucleic acid detection results of the BD MAX™ System and other platforms to formulate an optimized laboratory verification process. The re-examination of 400 samples determined as positive by BD MAX™ indicated that the inconsistency rate between BD MAX™ and the other platforms was 65.8%; the inconsistency rate of single-gene-positive results was as high as 99.2%. A receiver operating characteristic curve was drawn for the relative light unit (RLU) values of samples positive for a single gene, and RLU 800 was used as the cutoff. After setting the retest standard as single-gene positive and RLU ≥ 800, the number of the 260 BD MAX™ single-gene positives that needed to be confirmed again was 36 (13.8%) and the number that could be directly reported as negative was 224 (86.2%). This verification process can shorten the reporting period and speed up the epidemic adjustment time and turnover rate of special wards, thereby improving SARS-CoV-2 detection efficiency and clinical decision-making.