Journal of Arrhythmia (Apr 2018)

High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer

  • Tatsuhiko Hirao,
  • Junichi Nitta,
  • Akira Sato,
  • Yoshihide Takahashi,
  • Masahiko Goya,
  • Kenzo Hirao

DOI
https://doi.org/10.1002/joa3.12033
Journal volume & issue
Vol. 34, no. 2
pp. 198 – 200

Abstract

Read online

Abstract A 46‐year‐old female with a body mass index of 38.9 kg/m2 and no organic heart disease underwent a subcutaneous implantable cardioverter‐defibrillator implantation for secondary prevention of sudden cardiac death in the setting of idiopathic ventricular fibrillation. Defibrillation threshold (DFT) testing during implantation detected high shock impedance and high DFT. Fluoroscopy revealed subcoil fat between the lead and the sternum, which we suspected was the reason for the high shock impedance and high DFT. We repositioned the lead to a site just above the sternum and the shock impedance and DFT improved to within the respective normal ranges.