Biotechnology Reports (Jun 2020)

Taking advantage of a high-throughput flow cytometer for the implementation of an ADCC assay for regulatory compliance

  • Rosa Camacho-Sandoval,
  • Alexis Jiménez-Uribe,
  • Alejandra V. Tenorio-Calvo,
  • Carlos A. López-Morales,
  • Leslie Muñoz-García,
  • Alejandra Montes-Luna,
  • Héctor Leonardo García-Xolalpa,
  • Marco Velasco-Velázquez,
  • Lenin Pavón,
  • Sonia Mayra Pérez-Tapia,
  • Emilio Medina-Rivero

Journal volume & issue
Vol. 26
p. e00456

Abstract

Read online

Technological advances allowed the development of high-throughput instruments such as IntelliCyt iQue Screener PLUS®. Here, we took advantage of this technology to transfer a previously validated cytotoxicity assay. The evaluated parameters were cell permeability, caspase activation and phosphatidyl serine exposure. The assay was accurate (r2 = 0.90), precise (%CV ≤ 18.90) and specific. These results showed that this technology is suitable to be used in control quality environments. In addition, the automation provided a faster acquisition and analysis of data with precise and accurate results. This application could be implemented to evaluate another in vitro mechanism of action of different biotherapeutics.