Diagnostics (Jun 2024)
Validation of a New Test for Measuring the Contrast Sensitivity Function (Optopad-CSF) at Near Vision
Abstract
Our purpose is to develop and validate a new iPad-based contrast sensitivity (CS) test for measuring the contrast sensitivity function at near vision (Optopad-CSF). A total of 200 eyes of 100 healthy subjects (ages 17–63) were evaluated in a comparative study between the Optopad-CSF test (near vision) and the CSV-1000E test (distance vision). The agreement between tests was assessed with the index of contrast sensitivity (ICS) and the area under the curve (AUC). CS for all the spatial frequencies in both eyes showed a negative significant correlation with age, and corrected distance, and near visual acuities (r ≤ −0.512, p ≤ 0.013). A significantly lower CS was found with the Optopad-CSF test in the over-40-year-old subgroup for all the spatial frequencies evaluated compared to the below-40 subgroup (p ≤ 0.008). The mean AUC of the Optopad-CSF test (5.84) was twice that of the CSV-1000E test (2.76). The mean ICS of the Optopad-CSF (−0.019) and CSV-1000E (−0.075) tests showed similar values, both close to 0 (p = 0.3). There was a weak but significant correlation between the Optopad-CSF and CSV-1000E ICS tests (r = 0.246, p p < 0.001). Optopad-CSF is a valid portable system for measuring CS at near vision for five spatial frequencies, allowing the detection of age-related changes in CSF with age and CSF loss in cataracts, with no ceiling effect.