Sensors (Aug 2019)

Evaluation of Aerosol Electrospray Analysis of Metal-on-Metal Wear Particles from Simulated Total Joint Replacement

  • Gobert von Skrbensky,
  • Karoline Mühlbacher,
  • Emir Benca,
  • Alexander Kolb,
  • Reinhard Windhager,
  • Georg Reischl,
  • Georg Reinisch

DOI
https://doi.org/10.3390/s19173751
Journal volume & issue
Vol. 19, no. 17
p. 3751

Abstract

Read online

Wear is a common cause for aseptic loosening in artificial joints. The purpose of this study was to develop an automated diagnostical method for identification of the number and size distribution of wear debris. For this purpose, metal debris samples were extracted from a hip simulator and then analyzed by the electrospray method combined with a differential mobility analyzer, allowing particle detection ranging from several nanometers up to 1 µm. Wear particles were identified with a characteristic peak at 15 nm. The electrospray setup was successfully used and validated for the first time to characterize wear debris from simulated total joint replacement. The advantages of this diagnostic method are its time- and financial efficiency and its suitability for testing of different materials.