Scientific Reports (Aug 2023)

Iridocorneal angle imaging of a human donor eye by spectral-domain optical coherence tomography

  • Shangbang Luo,
  • Guy Holland,
  • Reza Khazaeinezhad,
  • Samantha Bradford,
  • Rohan Joshi,
  • Tibor Juhasz

DOI
https://doi.org/10.1038/s41598-023-37248-0
Journal volume & issue
Vol. 13, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Iridocorneal angle (ICA) details particularly the trabecular meshwork (TM), Schlemm’s canal (SC), and collector channels (CCs) play crucial roles in the regulation of the aqueous outflow in the eyes and are closely associated with glaucoma. Current clinical gonioscopy imaging provides no depth information, and studies of 3D high-resolution optical coherence tomography (OCT) imaging of these structures are limited. We developed a custom-built spectral-domain (SD-) OCT imaging system to fully characterize the angle details. Imaging of a human cadaver eye reveals the visibility of details in the TM/SC/CC region via a ’crossline’ scanning and a series of image processing. This shows that ICA imaging can be used for preoperative glaucoma inspections in the clinical setting with the proposed prototype.