EPJ Web of Conferences (Jan 2020)

High-count-rate 3He position-sensitive detector system: NEUNET-HCR

  • Sato Setsuo,
  • Seya Tomohiro,
  • Oshita Hidetoshi,
  • Kato Hideto,
  • Hikida Noriyuki,
  • Ishizawa Kazuya,
  • Yamaguchi Akinori,
  • Matsuura Mitsugu

DOI
https://doi.org/10.1051/epjconf/202023105004
Journal volume & issue
Vol. 231
p. 05004

Abstract

Read online

Highly reliable 3He gas detectors are most often used in neutron scattering experiments but have disadvantages such as low count rates and low position resolutions. In this study, a 3He gas detector system called NEUNET-HCR was developed, whose count rate is approximately 10 times that of the current NEUNET system, which was developed for a high position resolution with a low count rate. NEUNET-HCR achieved a maximum count rate of 535 kcps per detector, with a loss of approximately 30%. As the NEUNET-HCR system was developed based on the NEUNET system, various improvements were required to achieve the high count rate. Further, a high count rate was achieved in the NEUNET-HCR system without significantly degrading the position resolution.