Heliyon (May 2019)

Optimization of tris(2-carboxyethyl) phosphine reduction conditions for fast analysis of total biothiols in mouse serum samples

  • Chunfang Chang,
  • Muneki Isokawa,
  • Takashi Funatsu,
  • Makoto Tsunoda

Journal volume & issue
Vol. 5, no. 5
p. e01598

Abstract

Read online

In this study, we investigated suitable conditions for the reduction of disulfides in mouse serum samples by tris(2-carboxyethyl) phosphine (TCEP) for fast analysis of total biothiols. Disulfides were reduced with TCEP, and then, thiols were derivatized with the fluorogenic reagent, ammonium 7-fluoro-2,1,3-benzoxadiazole-4-sulfonate (SBD-F). Interference peaks on chromatograms of mouse serum samples disappeared when the TCEP reaction was conducted on ice instead of at room temperature, which is used classically. Low-molecular-weight disulfides, such as cystine and glutathione disulfide, were nearly completely reduced by TCEP on ice. Six SBD-biothiols (homocysteine, cysteine, cysteinylglycine, glutathione, γ-glutamylcysteine, and N-acetylcysteine) were separated within 7.5 min on a sulfoalkylbetain-type column (ZIC-HILIC: 150 × 2.1 mm i.d., 3.5 μm), without interference peaks. The developed method showed good linearity and reproducibility, with inter- and intra-day precisions of less than 3%.

Keywords