Biosensors (Mar 2023)

Microfluidic Electrochemical Glucose Biosensor with In Situ Enzyme Immobilization

  • Nina Lokar,
  • Borut Pečar,
  • Matej Možek,
  • Danilo Vrtačnik

DOI
https://doi.org/10.3390/bios13030364
Journal volume & issue
Vol. 13, no. 3
p. 364

Abstract

Read online

The development and characterization of a microfluidic electrochemical glucose biosensor are presented herein. The transducer part is based on thin-film metal electrodes on a glass substrate. The biological recognition element of the biosensor is the pyrroloquinoline quinone–glucose dehydrogenase (PQQ-GdhB) enzyme, selectively in situ immobilized via microcontact printing of a mixed self-assembling monolayer (SAM) on a gold working electrode, while the microfluidic part of the device comprises microchannel and microfluidic connections formed in a polydimethylsiloxane (PDMS) elastomer. The electrode properties throughout all steps of biosensor construction and the biosensor response to glucose concentration and analyte flow rate were characterized by cyclic voltammetry and chronoamperometry. A measurement range of up to 10 mM in glucose concentration with a linear range up to 200 μM was determined. A detection limit of 30 µM in glucose concentration was obtained. Respective biosensor sensitivities of 0.79 nA/µM/mm2 and 0.61 nA/µM/mm2 were estimated with and without a flow at 20 µL/min. The developed approach of in situ enzyme immobilization can find a wide number of applications in the development of microfluidic biosensors, offering a path towards continuous and time-independent detection.

Keywords