Journal of Analytical Methods in Chemistry (Jan 2012)

RP-HPLC Method Development and Its Validation for Quantitative Determination of Rimonabant in Human Plasma

  • Shravan Bankey,
  • Ganesh Tapadiya,
  • Jasvant Lamale,
  • Deepti Jain,
  • Shweta Saboo,
  • S. S. Khadabadi

DOI
https://doi.org/10.1155/2012/625979
Journal volume & issue
Vol. 2012

Abstract

Read online

A simple, accurate, and precise HPLC method was developed and validated for determination of rimonabant in human plasma. Following liquid-liquid extraction, chromatographic separation was accomplished using C18 column with mobile phase consisting of acetonitrile : water (90 : 10, v/v), drug was detected at 260 nm using UVdetector. The LOD and LOQ were 3.0 and 10.0 μg/L, respectively. The method is linear in the interval 50.0–1000.0 μg/L. The average extraction recovery of drug from plasma was found to be 92.2%. The percent CV of the method was found to be less than 10.8%, and accuracy was found between 94.5 and 106.7%. The assay may be applied to a pharmacokinetic and bioequivalence study of rimonabant.