Effect of long-term frozen plasma on the detection of HBsAg by ELISA

  • Simin LIAN,
  • Xiaohua CHEN,
  • Shubo LIN,
  • Ziyue GUO,
  • Zebin XIAO

DOI
https://doi.org/10.13303/j.cjbt.issn.1004-549x.2024.01.016
Journal volume & issue
Vol. 37, no. 1
pp. 91 – 94

Abstract

Read online

Objective To evaluate the HBsAg detection results of HBsAg+ samples after 8 years refrigeration by ELISA and evaluate the effectiveness of the current storage policy of retained samples. Methods A total of 100 HBsAg+ plasma samples by ELISA from May 2014 to March 2015 and refrigerated at -20℃ were collected and retested for HBsAg using the same method after thawing in 2023. Results The HBsAg retest results of 100 plasma samples were all positive, with the concordance rate of 100%, though there was a significant decrease in the S/CO value after refrigeration(27.52 vs 19.03, P<0.05). Conclusion Long-term refrigeration can lead to a decrease in the S/CO value of HBsAg ELISA detection,but it does not affect the positive results.