BMC Ophthalmology (Dec 2022)
Comparison of anterior corneal aberrations measured by Scheimpflug and Placido Disc System for myopes
Abstract
Abstract Background To ascertain the agreement of corneal aberrations obtained from the Pentacam and the KR-1W in myopic populations and to investigate the influence of the level of myopia as well as the laterality on the agreement. Methods In this observational study, a rotating Scheimpflug camera (Pentacam AXL) and a Hartmann-Shack wavefront analyzer with Placido-disc topographer (KR-1W) were used to measure the aberrations of myopes in the anterior corneal surface by one experienced operator. All examinations were computed across a 6 mm diameter. Six subgroups were generated according to the degree of myopia (mild, moderate, and severe myopia) and the laterality of eyes (right and left eyes). Results The study included 245 eyes of 170 participants. For certain anterior corneal aberrations, statistically significant differences existed between the Pentacam and the KR-1W (all P 0.1 μm) for aberrations closely correlated with visual quality, including Z(3, ± 3), Z(3, ± 1), and Z(4,0), and almost all aberrations, indicating poor agreement. Conclusions In clinical practice, the Pentacam based on Scheimpflug technology and the KR-1W based on Placido Disc System are not interchangeable in measuring anterior corneal aberration for myopes regardless of myopia degree and the laterality, suggesting that a consistent instrument should be selected for surgical design as well as follow-up.