JACC: Case Reports (Feb 2024)
Very Late Patent Foramen Ovale Occluder Device Structural Dysfunction and Thrombosis
Abstract
Mechanical dysfunction of patent foramen ovale (PFO) closure device is extremely rare. We present a 58-year-old male patient who had multiple episodes of ischemic strokes 3 years after PFO closure, which was related to PFO device mechanical dysfunction and thrombosis. He was successfully treated with surgical intervention.