Brazilian Journal of Radiation Sciences (Nov 2017)

Condições de radiação em tomografia computadorizada (TC): determinação e calibração de dosímetros

  • Lúcio Chagas Andrade,
  • José Guilherme Pereira Peixoto

DOI
https://doi.org/10.15392/bjrs.v5i3-A.299
Journal volume & issue
Vol. 5, no. 3-A

Abstract

Read online

RQT is the standard for radiation conditions in computed tomography. It simulates a beam unrelieved of a CT scanner . The camera pencil ionization dosimeter is used in CT. The LNMRI become known characterization of RQT radiation conditions and the secondary standard calibration for type pencil ionisation chambers in its lab. The obtained beam has the characteristics required by IEC 61267. The results of the calibration presented combined uncertainty expanded to 95.45% from 2.22%.