Scientific Reports (Oct 2024)
Comparison of a Scheimpflug imaging with other screening indices in diagnosing keratoconus and keratoconus suspect
Abstract
Abstract Keratoconus (KC) is an irreversible blinding eye disease; therefore, early screening of KC suspects (KCS) is crucial for protecting patients’ quality of life. Scheimpflug imaging is a commonly used screening device in clinical practice. We aimed to evaluate the diagnostic ability of a Scheimpflug imaging device (Scansys) for KC and KCS and compared it with other Scheimpflug-based devices (Pentacam and Corvis ST). This prospective case-control study included 107 normal eyes, 72 KCS, and 57 KC. Scansys screening index Keratoconus probability (KCP) showed excellent performance in diagnosing KC at a cutoff value of 16.4 (area under the receiver operating characteristic [AUROC] = 1.000), with 100% sensitivity and 98.11% specificity. KCP had a better KCS diagnostic ability at a cutoff value of 8.9 (AUROC = 0.813) than Corvis biomechanical index (CBI, AUROC = 0.764), reaching 67.61% sensitivity and 85.85% specificity. Pentacam screening index Belin/Ambrósio enhanced ectasia display deviation (BAD-D) showed the best performance with 92.96% sensitivity and 89.62% specificity at a cutoff value of 1.525 (AUROC = 0.970) in diagnosing KCS. Scansys provides accurate KCP parameters in diagnosing KC; however, the efficiency of diagnosing KCS should be further optimized.