Clinical and Experimental Otorhinolaryngology (Mar 2008)
Clinical Outcomes of Cochlear Reimplantation Due to Device Failure
Abstract
ObjectivesThe aim of this study was to evaluate the clinical features of cochlear reimplantation due to device failure.MethodsThe medical records of 30 patients who had undergone a revision cochlear implantation were retrospectively reviewed. Causes of revision operations, number of electrode channels inserted, and postoperative speech performances were analyzed.ResultsDevice failure (N=12, 38.7%) and hematoma (N=3, 9.6%) were the two most common reasons for revision surgery. In patients with device failure, the number of electrode channels reinserted was equal to, or more than the number of channels inserted during initial implantation. Speech performance scores remained the same, or improved after reimplantation in patients with device failure.ConclusionDevice failure was the most common cause of revision operation in patients with cochlear implanttion. Contrary to expectation, new electrodes were fully inserted without difficulty in all reimplantation cases. Intracochlear damage due to reimplantation appeared to be clinically insignificant.