Journal of Anaesthesiology Clinical Pharmacology (Jan 2013)

Carbon dioxide rebreathing caused by deformed silicon leaflet in the expiratory unidirectional valve

  • Arumugam Vasudevan,
  • Satyen Parida,
  • Lenin Babu Elakkumanan,
  • Sandeep Mishra

DOI
https://doi.org/10.4103/0970-9185.105816
Journal volume & issue
Vol. 29, no. 1
pp. 108 – 110

Abstract

Read online

Rebreathing of carbon dioxide caused by incompetent ′cage and disc′ unidirectional valves has been reported earlier. Some manufacturers have changed the design of unidirectional valves to ′flexible leaflets′. We report a series of cases where a deformed membrane leaflet in expiratory unidirectional valves led to rebreathing of carbon dioxide.

Keywords