Gels (Sep 2022)

Three-Dimensional Dosimetry by Optical-CT and Radiochromic Gel Dosimeter of a Multiple Isocenter Craniospinal Radiation Therapy Procedure

  • Matheus Antonio da Silveira,
  • Juliana Fernandes Pavoni,
  • Alexandre Colello Bruno,
  • Gustavo Viani Arruda,
  • Oswaldo Baffa

DOI
https://doi.org/10.3390/gels8090582
Journal volume & issue
Vol. 8, no. 9
p. 582

Abstract

Read online

Craniospinal irradiation (CSI) is a complex radiation technique employed to treat patients with primitive neuroectodermal tumors such as medulloblastoma or germinative brain tumors with the risk of leptomeningeal spread. In adults, this technique poses a technically challenging planning process because of the complex shape and length of the target volume. Thus, it requires multiple fields and different isocenters to guarantee the primary-tumor dose delivery. Recently, some authors have proposed the use IMRT technique for this planning with the possibility of overlapping adjacent fields. The high-dose delivery complexity demands three-dimensional dosimetry (3DD) to verify this irradiation procedure and motivated this study. We used an optical CT and a radiochromic Fricke-xylenol-orange gel with the addition of formaldehyde (FXO-f) to evaluate the doses delivered at the field junction region of this treatment. We found 96.91% as the mean passing rate using the gamma analysis with 3%/2 mm criteria at the junction region. However, the concentration of fail points in a determined region called attention to this evaluation, indicating the advantages of employing a 3DD technique in complex dose-distribution verifications.

Keywords