Diagnostics (Jan 2023)

VIS-NIR Diffuse Reflectance Spectroscopy System with Self-Calibrating Fiber-Optic Probe: Study of Perturbation Resistance

  • Valeriya Perekatova,
  • Alexey Kostyuk,
  • Mikhail Kirillin,
  • Ekaterina Sergeeva,
  • Daria Kurakina,
  • Olga Shemagina,
  • Anna Orlova,
  • Aleksandr Khilov,
  • Ilya Turchin

DOI
https://doi.org/10.3390/diagnostics13030457
Journal volume & issue
Vol. 13, no. 3
p. 457

Abstract

Read online

We report on the comparative analysis of self-calibrating and single-slope diffuse reflectance spectroscopy in resistance to different measurement perturbations. We developed an experimental setup for diffuse reflectance spectroscopy (DRS) in a wide VIS-NIR range with a fiber-optic probe equipped with two source and two detection fibers capable of providing measurements employing both single- and dual-slope (self-calibrating) approaches. In order to fit the dynamic range of a spectrometer in the wavelength range of 460–1030 nm, different exposure times have been applied for short (2 mm) and long (4 mm) source-detector distances. The stability of the self-calibrating and traditional single-slope approaches to instrumental perturbations were compared in phantom and in vivo studies on human palm, including attenuations in individual channels, fiber curving, and introducing optical inhomogeneities in the probe–tissue interface. The self-calibrating approach demonstrated high resistance to instrumental perturbations introduced in the source and detection channels, while the single-slope approach showed resistance only to perturbations introduced into the source channels.

Keywords