Journal of Optometry (Jul 2012)

Comparison of Keeler Pulsair EasyEye tonometer and Ocular Response Analyzer for measuring intraocular pressure in healthy eyes

  • Carmen Lopez-De La Fuente,
  • Ana Sanchez-Cano,
  • Antonio Ferreras,
  • Isabel Fuertes-Lazaro

DOI
https://doi.org/10.1016/j.optom.2012.06.003
Journal volume & issue
Vol. 5, no. 3
pp. 139 – 146

Abstract

Read online

Purpose: To evaluate the relationship between intraocular pressure (IOP) measurements obtained with Pulsair EasyEye (PEE) and Ocular Response Analyser (ORA) in healthy patients. Methods: Sixty-five eyes from 65 patients underwent a full optometric examination, including central corneal thickness (CCT), and IOP measured with PEE and ORA. Differences between IOP measurements between both tonometers were analyzed. Pearson correlation coefficients between IOP values and ORA corneal biomechanics parameters were also obtained. Results: Statistically significant differences were found between IOP of PEE (IOPk) and Goldmann-corrected IOP of ORA (IOPg; p = 0.001). IOPk and corneal resistance-corrected IOP of ORA (IOPcc) were also found to differ significantly (p = 0.025). Mean differences between IOPg-IOPk, IOPcc-IOPk and IOPg-IOPcc were 0.71 ± 1.66, 0.70 ± 2.46 and 0.01 ± 1.54 mmHg (mean ± standard deviation), respectively. Pearson correlation coefficients indicated that IOPk, IOPg, and IOPcc were significantly correlated among them (p < 0.001): r = 0.816 for IOPk-IOPg, r = 0.826 for IOPcc-IOPg and r = 0.587 IOPcc-IOPk. IOPk and IOPg were linearly associated with corneal resistance factor (CRF; r = 0.626 and r = 0.619, respectively) and with CCT (r = 0.531 and r = 0.579, respectively). IOPcc had a linear relationship with corneal hysteresis (CH) (r = −0.482) and similar results were found between CRF and CH (r = 0.841), CRF and CCT (r = 0.681) or between CH and CCT (r = 0.466). Conclusions: Differences between mean values of IOP measured with PEE and ORA are statistically significant, with ORA tonometer taking higher IOP values than PEE in most of the cases. IOPk, IOPcc and IOPg have, al least, moderate positive linear correlations and ORA biomechanics parameters CRF, CH and CCT have a linear positive relation between them.

Keywords