Applied Sciences (Jan 2019)

Ultrasonic Parametrization of Arterial Wall Movements in Low- and High-Risk CVD Subjects

  • Monika Makūnaitė,
  • Rytis Jurkonis,
  • Alberto Rodríguez-Martínez,
  • Rūta Jurgaitienė,
  • Vytenis Semaška,
  • Karolina Mėlinytė,
  • Raimondas Kubilius

DOI
https://doi.org/10.3390/app9030465
Journal volume & issue
Vol. 9, no. 3
p. 465

Abstract

Read online

This paper shows the results of a preliminary study on the performance of new methods based on ultrasonic images parametrization, to estimate the arterial wall movements used for the evaluation of arterial stiffness, considered to be a predictor of cardiovascular events. The well-known technique of motion tracking in ultrasound image sequences was applied on cine loops scanned from subjects with different risks of suffering from cardiovascular disease (CVD). The motion of arterial walls was traced using displacement signals: Diameter, intima-media thickness (IMT) and longitudinal intima-media (IM) complex movement. The new methods used for the parametrization of the displacement signals were the average value (AV), effective or root mean square (RMS) value, and peak-to-peak motion amplitude estimate. A total of 79 subjects were analyzed in the study with 30 considered at low risk and 49 included in a preventive program for monitoring high CVD risk subjects. The results show a statistically significant difference between healthy volunteers and at-risk patients according to the AV of IMT, RMS values of longitudinal and radial motions and peak-to-peak amplitude of radial motion.

Keywords