Journal of Investigative Surgery (Jan 2022)
Comparison of Centrifugal and Pulsatile Perfusion to Preserve Donor Kidneys Using Ex Vivo Subnormothermic Perfusion
Abstract
Objective We have previously demonstrated benefits of kidney preservation utilizing an oxygenated subnormothermic ex vivo perfusion platform. Herein, we aim to compare pulsatile versus centrifugal (steady and uniform flow) perfusion with the goal of optimizing renal preservation with these devices. Materials and methods: Pig kidneys were procured following 30 min of warm ischemia by cross-clamping both renal arteries. Paired kidneys were cannulated and underwent either: oxygenated pulsatile or centrifugal perfusion using a hemoglobin oxygen carrier at room temperature with our ex vivo machine perfusion platform for 4 hr. Kidneys were reperfused with whole blood for 4 hr at 37° C. Renal function, pathology and evidence of inflammation were assessed post-perfusion. Results: Both pump systems performed equally well with organs exhibiting similar renal blood flow, and function post-reperfusion. Histologic evidence of renal damage using apoptosis staining and acute tubular necrosis scores was similar between groups. This was corroborated with urinary assessment of renal damage (NGAL 1) and inflammation (IL-6), as levels were similar between groups. Conclusion: In our porcine model with added warm ischemia simulating the effects of reperfusion after transplantation, pulsatile perfusion yielded similar renal protection compared with centrifugal perfusion kidney preservation. Both methods of perfusion can be used in ex vivo kidney perfusion systems.
Keywords