Diagnostic and Interventional Radiology (Mar 2019)

Virtual touch tissue imaging quantification shear wave elastography for determining benign versus malignant cervical lymph nodes: a comparison with conventional ultrasound

  • Alican Kılıç,
  • Hale Çolakoğlu Er

DOI
https://doi.org/10.5152/dir.2019.18406
Journal volume & issue
Vol. 25, no. 2
pp. 114 – 121

Abstract

Read online

PURPOSEWe aimed to prospectively examine virtual touch tissue imaging quantification (VTIQ) shear wave elastography (SWE) and conventional (B-mode and Doppler) ultrasonography (US), individually and combined, for their ability to differentiate benign and malignant cervical lymph nodes (CLNs).METHODSOne hundred enlarged lymph nodes (LNs) from 72 patients, confirmed by histopathologic diagnoses, were included in the present study. B-mode US, Doppler US, and SWE were performed before histopathologic sampling of the LNs. The LN shear wave velocity (SWV, m/s) was assessed by VTIQ.RESULTSUsing a 3.03 m/s cutoff value, the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of SWVmax for differentiating benign and malignant CLNs were 93%, 59%, 68%, 91%, and 75%, respectively. B-mode and Doppler had a sensitivity, specificity, PPV, NPV, and accuracy of 79.2%, 71.2%, 71.6%, 78.7%, and 75%, respectively. B-mode, Doppler, and VTIQ SWE combined had 87.5%, 75%, 76.3%, 86.6%, and 81% for the same parameters, respectively.CONCLUSIONVTIQ SWE is a promising noninvasive diagnostic imaging technique for differentiating benign and malignant CLNs. VTIQ SWV can improve the diagnostic performance of conventional US for differentiating benign and malignant CLNs.