Scientific Reports (Feb 2021)

The design of X-band EPR cavity with narrow detection aperture for in vivo fingernail dosimetry after accidental exposure to ionizing radiation

  • Junwang Guo,
  • Xiaoxiao Luan,
  • Ye Tian,
  • Lei Ma,
  • Xiaoguang Bi,
  • Jierui Zou,
  • Guofu Dong,
  • Ye Liu,
  • Yonggang Li,
  • Jing Ning,
  • Ke Wu

DOI
https://doi.org/10.1038/s41598-021-82462-3
Journal volume & issue
Vol. 11, no. 1
pp. 1 – 8

Abstract

Read online

Abstract For the purpose of assessing the radiation dose of the victims involved in the nuclear emergency or radiation accident, a new type of X-band EPR resonant cavity for in vivo fingernail EPR dosimetry was designed and a homemade EPR spectrometer for in vivo fingernail detection was constructed. The microwave resonant mode of the cavity was rectangular TE101, and there was a narrow aperture for fingernail detection opened on the cavity’s wall at the position of high detection sensitivity. The DPPH dot sample and the fingernail samples were measured based on the in vivo fingernail EPR spectrometer. The measurements of the DPPH dot sample verified the preliminary functional applicable of the EPR spectrometer and illustrated the microwave power and modulation response features. The fingernails after irradiation by gamma-ray were measured and the radiation-induced signal was acquired. The results indicated that the cavity and the in vivo EPR dosimeter instrument was able to detect the radiation-induced signal in irradiated fingernail, and preliminarily verified the basic function of the instrument and its potential for emergency dose estimate after a radiation accident.