Biomedicines (Mar 2023)

Proving the Superiority of Intraoperative Recurrent Laryngeal Nerve Monitoring over Visualization Alone during Thyroidectomy

  • Beata Wojtczak,
  • Dominik Marciniak,
  • Krzysztof Kaliszewski,
  • Krzysztof Sutkowski,
  • Mateusz Głód,
  • Jerzy Rudnicki,
  • Marek Bolanowski,
  • Marcin Barczyński

DOI
https://doi.org/10.3390/biomedicines11030880
Journal volume & issue
Vol. 11, no. 3
p. 880

Abstract

Read online

Vocal fold paralysis after thyroid surgery is still a dangerous complication that significantly reduces patients’ quality of life. Since the intraoperative neuromonitoring (IONM) technique has been introduced and standardized, the most frequently asked question is whether its use has significantly reduced the rate of RLN injury during thyroid surgery compared to visual identification alone (VA). The aim of this study was to attempt to prove the superiority of IONM over VA of the RLN during thyroid surgery in the prevention of vocal fold paralysis, taking into account risk factors for complications. The medical records of 711 patients (1265 recurrent laryngeal nerves at risk of injury) were analyzed retrospectively: in 257 patients/469 RLNs at risk, thyroid surgery was performed with IONM; in 454 patients/796 RLNs at risk, surgery was performed with VA. The statistical analysis showed that in the group of patients with IONM only one risk factor—the surgeon’s experience—proved statistically significant (OR = 3.27; p = 0.0478) regarding the overall risk of vocal fold palsy. In the group of patients where only visualization was used, 5 of the 12 factors analyzed were statistically significant: retrosternal goiter (OR = 2.23; p = 0.041); total thyroid volume (OR = 2.30; p = 0.0284); clinical diagnosis (OR = 2.5; p = 0.0669); gender (OR = 3.08; p = 0.0054) and risk stratification (OR = 3.30; p = 0.0041). In addition, the cumulative risk, taking into account the simultaneous influence of all 12 factors, was slightly higher in the group of patients in whom only VA was used during the procedure: OR = 1.78. This value was also considerably more statistically significant (p p = 0.004. Conclusions: Risk factors for complications in thyroid surgery are not significant for any increase in the rate of vocal fold paralysis as long as surgery is performed with IONM, in contrast to thyroid surgery performed only with VA, thus proving the superiority of IONM over VA for safety.

Keywords