BMC Musculoskeletal Disorders (Aug 2022)

Finite element analysis after rod fracture of the spinal hybrid elastic rod system

  • Jui-Yang Hsieh,
  • Chen-Sheng Chen,
  • Shao-Ming Chuang,
  • Jyh-Horng Wang,
  • Po-Quang Chen,
  • Yi-You Huang

DOI
https://doi.org/10.1186/s12891-022-05768-x
Journal volume & issue
Vol. 23, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Background The spinal hybrid elastic (SHE) rod dynamic stabilization system can provide sufficient spine support and less adjacent segment stress. This study aimed to investigate the biomechanical effects after the internal fracture of SHE rods using finite element analysis. Methods A three-dimensional nonlinear finite element model was developed. The SHE rod comprises an inner nitinol stick (NS) and an outer polycarbonate urethane (PCU) shell (PS). The fracture was set at the caudal third portion of the NS, where the maximum stress occurred. The resultant intervertebral range of motion (ROM), intervertebral disc stress, facet joint contact force, screw stress, NS stress, and PCU stress were analyzed. Results When compared with the intact spine model, the overall trend was that the ROM, intervertebral disc stress, and facet joint force decreased in the implanted level and increased in the adjacent level. When compared with the Ns-I, the trend in the Ns-F decreased and remained nearly half effect. Except for torsion, the PS stress of the Ns-F increased because of the sharing of NS stress after the NS fracture. Conclusions The study concluded the biomechanical effects still afford nearly sufficient spine support and gentle adjacent segment stress after rod fracture in a worst-case scenario of the thinnest PS of the SHE rod system.