Biosensors (Jan 2023)

The Integration of Reference Electrode for ISFET Ion Sensors Using Fluorothiophenol-Treated rGO

  • Dae Hoon Kim,
  • Hae Shin Cho,
  • Jin Heung Kim,
  • Da Ae Jo,
  • Hong Gi Oh,
  • Byoung Kuk Jang,
  • Kwang Soup Song

DOI
https://doi.org/10.3390/bios13010089
Journal volume & issue
Vol. 13, no. 1
p. 89

Abstract

Read online

Ion-sensitive field-effect transistors (ISFETs) detect specific ions in solutions that enable straightforward, fast, and inexpensive sensors compared to other benchtop equipment. However, a conventional reference electrode (RE) such as Ag/AgCl is limited on the miniaturization of the sensor. We introduce reduced graphene oxide (rGO), which serves as a new RE, when fluorinated (F-rGO) using fluorothiophenol through the π–π interaction. The circular RE is integrated between a fabricated microscale two-channel ISFET, which is capable of detecting two kinds of ions on an indium tin oxide (ITO) thin-film substrate, using the photolithography process. F-rGO bound to this circular region to function as an RE in the ISFETs sensor, which operated stably in solution and showed a relatively high transconductance (gm) value (1.27 mS), low drift characteristic (3.2 mV), and low hysteresis voltage (±0.05 mV). It detected proton (H+) ions in a buffer solution with high sensitivity (67.1 mV/pH). We successfully detected Na+ (62.1 mV/dec) and K+ (57.6 mV/dec) ions in human patient urine using a two-channel ISFET with the F-rGO RE. The F-rGO RE will be a suitable component in the fabrication of low-cost, mass-produced, and disposable ISFETs sensors.