Journal of Cardiothoracic Surgery (Mar 2019)

A case report on epicardial ultrasonography of coronary anastomoses using a stabilizing device without the use of ultrasound gel

  • Jan Jesper Andreasen,
  • Dorte Nøhr,
  • Alex Skovsbo Jørgensen

DOI
https://doi.org/10.1186/s13019-019-0882-2
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 5

Abstract

Read online

Abstract Background Intraoperative epicardial ultrasonography of coronary artery bypass graft anastomoses is a procedure used for anatomical quality assessment of peripheral anastomoses during coronary artery bypass grafting. However, it may be difficult to keep the ultrasound transducer in steady contact with the anastomoses on the beating heart without causing any deformation. Furthermore, we are not aware of any sterile ultrasound gel approved for application into the pericardial space. Case presentation We report a method using a stabilizing connecting device for an ultrasound transducer to be used for visualization of coronary anastomoses without application of ultrasound gel during on-pump coronary bypass surgery. Conclusion Use of a stabilizing device and coagulated blood from the patient as an alternative for ultrasound gel facilitates peroperative ultrasonography of coronary anastomoses. The procedure provides surgeons with non-deformed echocardiographic longitudinal and transverse images of all parts of the anastomoses. Trial registration The patient participated in a still ongoing clinical feasibility study: Trial registration: ClinicalTrials.gov ID: NCT02919124; Registered September 29, 2016.