Journal of NeuroEngineering and Rehabilitation (Jan 2021)

Symmetry function in gait pattern analysis in patients after unilateral transfemoral amputation using a mechanical or microprocessor prosthetic knee

  • Mateusz Kowal,
  • Sławomir Winiarski,
  • Ewa Gieysztor,
  • Anna Kołcz,
  • Karolina Walewicz,
  • Wojciech Borowicz,
  • Alicja Rutkowska-Kucharska,
  • Małgorzata Paprocka-Borowicz

DOI
https://doi.org/10.1186/s12984-021-00810-w
Journal volume & issue
Vol. 18, no. 1
pp. 1 – 12

Abstract

Read online

Abstract Background Above-knee amputations (AKAs) contribute to gait asymmetry. The level of asymmetry is affected by the type of knee prosthetic module. There is limited evidence suggesting that more technically advanced solutions (microprocessor modules; MicPK) are superior to less advanced ones (mechanical modules; MechPK). The study aimed to evaluate the variable range of hip and pelvic joint movements during gait and look for differentiating areas with an increased level of asymmetry of the gait cycle in individuals who underwent an AKA and are equipped with MicPK or MechPK. Methods Twenty-eight individuals participated in the study; 14 were assigned to a study group of individuals who underwent a unilateral AKA, and the other 14 were healthy participants as a control group. The movement task was recorded using the optoelectronic SMART-E system following the standard Davis protocol (the Newington model). A new method of quantifying gait symmetry using the symmetry function (SF) is proposed. SF is an integral measure of absolute differences in time-standardized signals between sides throughout the whole cycle of motion variability. Results In the frontal plane, there were significant differences between groups in the asymmetry of the range of movement in the hip joint of the intact limb. In the middle of the support phase, the intact limb was more adducted in individuals with MicPK and less abducted in people with MechPK (differences in mean 180%, p < 0.000; max 63%, p < 0.000; min 65%, p < 0.000). In the sagittal plane, the range of asymmetry of the flexion and thigh extension of the intact limb was similar to and only slightly different from the physiological gait. In the transverse plane, higher asymmetry values were noted for individuals with MicPK. In the final stage of the swing phase, the thigh was more rotated both externally and internally. The size of the asymmetry, when compared to gait of healthy individuals, reached 50% (differences in mean 115%, p < 0.232; max 62% p < 0.26; min 50, p < 0.154). Conclusions In the study group, the assessed ranges of pelvic and thigh movement in the hip joint differed only in the frontal plane. Individuals who underwent a unilateral above-knee amputation may gain less from using MicPK than anticipated.

Keywords