International Journal of Vascular Medicine (Jan 2013)

Transcranial Doppler Ultrasound: A Review of the Physical Principles and Major Applications in Critical Care

  • Jawad Naqvi,
  • Kok Hooi Yap,
  • Gulraiz Ahmad,
  • Jonathan Ghosh

DOI
https://doi.org/10.1155/2013/629378
Journal volume & issue
Vol. 2013

Abstract

Read online

Transcranial Doppler (TCD) is a noninvasive ultrasound (US) study used to measure cerebral blood flow velocity (CBF-V) in the major intracranial arteries. It involves use of low-frequency (≤2 MHz) US waves to insonate the basal cerebral arteries through relatively thin bone windows. TCD allows dynamic monitoring of CBF-V and vessel pulsatility, with a high temporal resolution. It is relatively inexpensive, repeatable, and portable. However, the performance of TCD is highly operator dependent and can be difficult, with approximately 10–20% of patients having inadequate transtemporal acoustic windows. Current applications of TCD include vasospasm in sickle cell disease, subarachnoid haemorrhage (SAH), and intra- and extracranial arterial stenosis and occlusion. TCD is also used in brain stem death, head injury, raised intracranial pressure (ICP), intraoperative monitoring, cerebral microembolism, and autoregulatory testing.