Practical Laboratory Medicine (Mar 2024)

Detection of infliximab, adalimumab, and anti-drug antibodies: Development and validation of new monotest, automated assays on multiparametric instrument

  • Helena Cerutti,
  • Giulia Tesi,
  • Francesco Petrini,
  • Tommaso Bandini,
  • Alessandra Cartocci,
  • Andrea Ianniello,
  • Alessia Bogi,
  • Chiara Muzzi,
  • Alessandra Brogi

Journal volume & issue
Vol. 39
p. e00374

Abstract

Read online

Objective: To convert manual ELISA kits to fully automated immunoassays that quantify serum drug levels and anti-drug antibodies levels of infliximab and adalimumab (CHORUS Promonitor kits). Desing and methods: CHORUS Promonitor INFLIXIMAB, CHORUS Promonitor ADALIMUMAB, CHORUS Promonitor ANTI-INFLIXIMAB, and CHORUS Promonitor ANTI-ADALIMUMAB kits were compared with the corresponding Promonitor kits to determine sensitivity and specificity of the assays. For the automated assays, the entire procedure -from samples dilution to final readouts-was performed by CHORUS TRIO instrument (DIESSE, Italy). Residual human serum samples from clinical laboratory investigations and samples resulting from the addition of specific drugs (IFX or ADL) or anti-drug antibodies (anti-IFX or anti-ADL) were used for the characterization and validation of the tests. Results: CHORUS Promonitor kits showed an excellent agreement (Cohen's coefficient = 1) with the Promonitor kits and were linear within predefined ranges. All assays were accurate and repeatable, as an acceptable variability were observed within runs, between runs, lots, and instruments. No difference in detecting the reference drug or biosimilars emerged. Conclusion: During preclinical development, these kits resulted as sensitive, specific, accurate, and able to quantify either the reference drug or the corresponding biosimilars. All these features support their use in clinical practice for therapeutic drug monitoring of patients with inflammatory diseases under treatment with IFX or ADL.

Keywords