Journal of International Medical Research (Jun 2020)

Feasibility of using calibrated cone-beam computed tomography scans to validate the heart dose in left breast post-mastectomy radiotherapy

  • Bin Tang,
  • Jiabao Ma,
  • Jinghui Xu,
  • Jie Li,
  • Shengwei Kang,
  • Pei Wang,
  • Fan Wu,
  • Lucia Clara Orlandini

DOI
https://doi.org/10.1177/0300060520929168
Journal volume & issue
Vol. 48

Abstract

Read online

Objective In post-mastectomy radiotherapy, high-conformal techniques are a valid method for determining the dose distribution around a target. However, the proximity of critical structures is a reason for concern. This study aims to evaluate the feasibility of using calibrated cone-beam computed tomography (CBCT) scans as a valid tool for a timely heart dose evaluation. Methods A retrospective analysis was conducted on 170 retrospective CBCT scans of 17 patients who underwent high-conformal post-mastectomy irradiation. The delivered doses that were calculated using personalized calibrated CBCT were compared with the doses planned, using the dose–volume histogram dosimetric parameters. Results The heart volume that was evaluated using CBCT presented a mean increase of 6%; this discrepancy impacted the heart dose in 4 of 17 patients, with an absolute increase of V25 Gy (range, 2.5%–7.6%) and an increase in the mean dose (range, 1.1–3.4 Gy). The dose for the target, ipsilateral lung, and contralateral breast remained unchanged. Conclusion Using CBCT to monitor the dose that is delivered to the heart is feasible, allowing for a timely shift to an adaptive plan if clinically necessary.