International Journal of Ophthalmology (Feb 2024)

Two different autorefractors for vision screening in children and adolescents

  • Xiao-Jun Chen,
  • Lin-Jie Liu,
  • Bing Sun,
  • Dan-Dan Jiang,
  • Shu-Shu Zuo,
  • Yan-Hui Wang,
  • Chun-Hua Zhang,
  • Yan-Yan Chen

DOI
https://doi.org/10.18240/ijo.2024.02.16
Journal volume & issue
Vol. 17, no. 2
pp. 331 – 338

Abstract

Read online

AIM: To compare the consistency of two autorefractors (Tianle RM-9000 and Topcon KR-800) for school-age myopia children, and to provide a basis for largescale data analysis and comparison. METHODS: The refractive error in 909 subjects (age 4–18y) were measured using both autorefractors without cycloplegia. The data were analyzed using Fourier decomposition and the correlation coefficients, intraclass correlation coefficients (ICC), and Bland-Altman limits of agreement (LoA) for each parameter were calculated. RESULTS: There was a strong correlation between the spherical equivalent (SE), sphere diopter (DS), and cylinder diopter (DC) readings of the Tianle RM-9000 and those of the Topcon KR-800, with correlation coefficient values of 0.98, 0.98 and 0.83 and ICC values of 0.99, 0.99 and 0.93, respectively. However, the correlation coefficients and ICC values of J0 and J45 were unreliable (R=-0.004, -0.034; both ICC<0.10). Bland-Altman analysis revealed that SE, DS, and DC measured by the Tianle RM-9000 were significantly biased toward myopia compared with the Topcon KR-800, and the mean differences were -0.072, -0.026, -0.091 D, respectively (all P<0.01). The minimum absolute value of the difference within the 95% LoA for SE, DS, and DC was 0.63 D, 0.50 D, 0.62 D, respectively; all these values were in the clinically acceptable range. For J0 and J45, the mean differences were close to zero (P=0.43, 0.84); however, the 95% LoA were relatively wide (J0 SD: 0.53; 95%CI: -1.00, 1.10; J45 SD: 0.52; 95%CI: -1.00, 1.00). CONCLUSION: The two autorefractors are consistent with each other, as the differences in SE, DS, and DC were within the clinically acceptable range. Readers can compare the data measured by either device in different studies and use the two devices in the same study to generate a dataset that can be analyzed together. However, the J0 and J45 vectors are unreliable and should not be used to assess astigmatism.

Keywords