Applied Sciences (Aug 2022)

The Role of a DirectDensity<sup>®</sup> CT Reconstruction in a Radiotherapy Workflow: A Phantom Study

  • Giacomo Feliciani,
  • Claretta Guidi,
  • Maria Luisa Belli,
  • Vincenzo D’Errico,
  • Emiliano Loi,
  • Emilio Mezzenga,
  • Anna Sarnelli

DOI
https://doi.org/10.3390/app12157845
Journal volume & issue
Vol. 12, no. 15
p. 7845

Abstract

Read online

The DirectDensity® CT reconstruction algorithm provides a reconstruction approach independent of the tube voltage, directly reconstructing the CT projection data into CT numbers related to the electron densities of the materials. This work examines the efficacy of DirectDensity® in the treatment planning process with both tissues and metallic materials. CT scans of a Cheese phantom were acquired at 80, 100, 120 and 140 kVp and reconstructed with different algorithms. Calibration curves were built for each kVp and reconstruction technique. To evaluate the flexibility of the DirectDensity® in dose calculations, a prostate cancer treatment plan was simulated on phantom images with and without metal inserts. Moreover, the robustness of the algorithm was tested by simulating a possible error in the selection of the calibration curve. As expected, the calibration curves related to DirectDensity® showed a tube voltage dependence only for densities above 1.82 g/cm3. The maximum percentage differences in dose distributions comparations never exceeded the 3% of tolerance and the 3D gamma analysis always returned indices greater than 90%. The results suggest that the DD reconstruction algorithm can be employed in most clinical cases and allows for a personalized radiotherapy cancer treatment workflow, maintaining its robustness and simplicity.

Keywords