Validation and application of a blood screening method for SARS-CoV-2 nucleic acid
Abstract
Objective To evaluate the performance and clinical application of a high-throughput nucleic acid blood screening detection system for SARS-CoV-2, so as to provide basis and technical means for its application in detection of plasma from recovered COVID-19 patients. Methods The SARS-CoV-2 nucleic acid was detected by real-time fluorescence quantitative PCR, and the sensitivity, precision, anti-interference and other parameters were evaluated. Blood donor samples collected during COVID-19 epidemic period were screened using the detection system to evaluate its applicability. Results The detection limits of gene N and ORF 1ab were 3.98 copies/mL and 9.38 copies/mL, respectively. The CV of high and low concentration samples were both less than 5%. Hemoglobin at 500 mg/dL and triglyceride at 3g/dL had little effect on the results. The detection system can effectively prevent carryover, thus avoiding false positive results. The SARS-CoV-2 nucleic acid blood screening was carried out in a total of 39 306 blood samples, and all samples were negative. Conclusion The established method can meet the needs of SARS-CoV-2 nucleic acid screening therefore ensure the safe application of plasma from recovered COVID-19 patients.