Current Directions in Biomedical Engineering (Sep 2020)
Guided capture of 3-D Ultrasound data and semiautomatic navigation using a mechatronic support arm system
Abstract
A common treatment for malignant hepatic tumors is radiofrequency thermal ablation (RFA); however this procedure is often exhausting and error-prone, when performed only with hand held devices. In this work, we present a new concept for a computer- and mechatronic-arm-assisted treatment of hepatic tumor with RFA. Challenging factors of state-of-the-art RFA procedures are tackled by utilizing a prototype, which combines state-of-the-art ultrasound image guidance with a mechatronic support arm system. Physicians can image and examine the human abdomen using an ultrasound capturing device which is guided by the mechatronic support arm. The mechatronic arm allows to track the ultrasound transducer via joint position encoders, so that the recorded images’ exact positions are acquired and the patient is automatically registered during imaging. Consequently, physicians can use the acquired ultrasound images to navigate in the imaged abdomen area. The established prototype enables preplanning and semi-automatic alignment of trajectories with sufficient accuracy. In addition, the arm system supports the usually lengthy, exhausting and error-prone holding of the needle during the ablation process. Based on the results of accuracy examination and expert assessments by surgeons, the established concept could be proven to be advantageous for the considered use case.
Keywords