Journal of Pharmaceutical Analysis (Oct 2015)

High-sensitivity simultaneous liquid chromatography–tandem mass spectrometry assay of ethinyl estradiol and levonorgestrel in human plasma

  • Abhishek Gandhi,
  • Swati Guttikar,
  • Priti Trivedi

Journal volume & issue
Vol. 5, no. 5
pp. 316 – 326

Abstract

Read online

A sensitive and simultaneous liquid chromatography–tandem mass spectrometry method was developed and validated for quantification of ethinyl estradiol and levonorgestrel. The analytes were extracted with methyl-tert-butyl ether: n-hexane (50:50, v/v) solvent mixture, followed by dansyl derivatization. The chromatographic separation was performed on a Kinetex C18 (50 mm×4.6 mm, 2.6 µm) column with a mobile phase of 0.1% (v/v) formic acid in water and acetonitrile in gradient composition. The mass transitions were monitored in electrospray positive ionization mode. The assay exhibited a linear range of 0.100–20.0 ng/mL for levonorgestrel and 4.00–500 pg/mL for ethinyl estradiol in human plasma. A run time of 9.0 min for each sample made it possible to analyze a throughput of more than 100 samples per day. The validated method has been successfully used to analyze human plasma samples for application in pharmacokinetic and bioequivalence studies. Keywords: Ethinyl estradiol, Levonorgestrel, LC–MS/MS, Human plasma, Derivatization