Intraoperative Beat-to-Beat Pulse Transit Time (PTT) Monitoring via Non-Invasive Piezoelectric/Piezocapacitive Peripheral Sensors Can Predict Changes in Invasively Acquired Blood Pressure in High-Risk Surgical Patients
Michael Nordine,
Marius Pille,
Jan Kraemer,
Christian Berger,
Philipp Brandhorst,
Philipp Kaeferstein,
Roland Kopetsch,
Niels Wessel,
Ralf Felix Trauzeddel,
Sascha Treskatsch
Affiliations
Michael Nordine
Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité–Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany
Marius Pille
Berlin Institute of Health at Charité, Universitätsmedizin Berlin, Charitéplatz 1, 10117 Berlin, Germany
Jan Kraemer
Department of Physics, Humboldt University zu Berlin, 10115 Berlin, Germany
Christian Berger
Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité–Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany
Philipp Brandhorst
Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité–Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany
Department of Physics, Humboldt University zu Berlin, 10115 Berlin, Germany
Ralf Felix Trauzeddel
Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité–Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany
Sascha Treskatsch
Department of Anesthesiology and Intensive Care Medicine, Hindenburgdamm 30, Charité–Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin and Humboldt Universität zu Berlin, 12203 Berlin, Germany
Background: Non-invasive tracking of beat-to-beat pulse transit time (PTT) via piezoelectric/piezocapacitive sensors (PES/PCS) may expand perioperative hemodynamic monitoring. This study evaluated the ability for PTT via PES/PCS to correlate with systolic, diastolic, and mean invasive blood pressure (SBPIBP, DBPIBP, and MAPIBP, respectively) and to detect SBPIBP fluctuations. Methods: PES/PCS and IBP measurements were performed in 20 patients undergoing abdominal, urological, and cardiac surgery. A Pearson’s correlation analysis (r) between 1/PTT and IBP was performed. The predictive ability of 1/PTT with changes in SBPIBP was determined by area under the curve (reported as AUC, sensitivity, specificity). Results: Significant correlations between 1/PTT and SBPIBP were found for PES (r = 0.64) and PCS (r = 0.55) (p IBP/DBPIBP for PES (r = 0.6/0.55) and PCS (r = 0.5/0.45) (p PES predicted a 30% SBPIBP decrease (0.82, 0.76, 0.76), while a 5.6% increase predicted a 30% SBPIBP increase (0.75, 0.7, 0.68). A 6.6% decrease in 1/PTTPCS detected a 30% SBPIBP decrease (0.81, 0.72, 0.8), while a 4.8% 1/PTTPCS increase detected a 30% SBPIBP increase (0.73, 0.64, 0.68). Conclusions: Non-invasive beat-to-beat PTT via PES/PCS demonstrated significant correlations with IBP and detected significant changes in SBPIBP. Thus, PES/PCS as a novel sensor technology may augment intraoperative hemodynamic monitoring during major surgery.