An observational, repeated measures design was used in this study to assess the validity of the Ergotex Inertial Measurement Unit (IMU) against a 3D motion capture system for measuring trunk, hip, and shoulder angles in ten healthy adult males (38.8 ± 7.3 y, bodyweight 79.2 ± 115.9 kg, body height 179.1 ± 8.1 cm). There were minimal systematic differences between the devices, with the most significant discrepancy being 1.4 degrees for the 80-degree target angle, denoting Ergotex’s precision in joint angle measurements. These results were statistically significant (p < 0.001), with predominantly trivial to small effect sizes, indicating high accuracy for clinical and biomechanical applications. Bland–Altman analysis showed Limits of Agreement (LoA) approximately ±2.5 degrees across all angles and positions, with overall LoA ranging from 3.6 to −2.4 degrees, reflecting Ergotex’s consistent performance. Regression analysis indicated uniform variance across measurements, with minor heteroscedastic errors producing a negligible underestimation trend of around 0.5 degrees in some instances. In conclusion, the Ergotex IMU is a reliable tool for accurate joint angle measurements. It offers a practical and cost-effective alternative to more complex systems, particularly in settings where precise measurement is essential.