Audiology Research (Sep 2021)
Investigating Performance of cVEMP and oVEMP in the Identification of Superior Canal Dehiscence in Relation to Dehiscence Location and Size
Abstract
Compare the sensitivity and specificity of cVEMP (500 Hz), oVEMP (500 Hz and 4 kHz) in the identification of SSCD. A secondary objective was to identify the influence of dehiscence size and location on cVEMP and oVEMP responses. Methods: Individuals with unilateral (n = 16) and bilateral (n = 10) scan confirmed SSCD were assessed using air-conducted cVEMP and oVEMP Results: For cVEMP, an amplitude cutoff of 286.9 μV or a threshold cutoff of 67.5 dBnHL revealed, respectively, a sensitivity of 75% and 70.6% and a specificity of 69.4% and 100%. For oVEMP (500 Hz), an amplitude cutoff of 10.8 μV or a threshold cutoff of 77.5 dBnHL revealed a sensitivity of 83.33% and a specificity of 87.5% and 80%, respectively. oVEMP (4 kHz), an amplitude cutoff of 3.1 μV, revealed a high specificity of 100% but a low sensitivity of 47.2%. A positive correlation was noted between the length of the SSCD and the cVEMP and oVEMP (500 Hz) thresholds and cVEMP amplitude. Conclusions: Our results support the use of oVEMP in the identification of SSCD. The presence of oVEMP (500 Hz) with an amplitude higher or equal to 10.8 μV, a threshold lower or equal to 77.5 dBnHL or oVEMP (4 kHz) amplitude of 3.1 μV represents the most useful to identify SSCD.