Audiology Research (Sep 2021)

Investigating Performance of cVEMP and oVEMP in the Identification of Superior Canal Dehiscence in Relation to Dehiscence Location and Size

  • Maxime Maheu,
  • Ahlem Elblidi,
  • Issam Saliba

DOI
https://doi.org/10.3390/audiolres11030042
Journal volume & issue
Vol. 11, no. 3
pp. 452 – 462

Abstract

Read online

Compare the sensitivity and specificity of cVEMP (500 Hz), oVEMP (500 Hz and 4 kHz) in the identification of SSCD. A secondary objective was to identify the influence of dehiscence size and location on cVEMP and oVEMP responses. Methods: Individuals with unilateral (n = 16) and bilateral (n = 10) scan confirmed SSCD were assessed using air-conducted cVEMP and oVEMP Results: For cVEMP, an amplitude cutoff of 286.9 μV or a threshold cutoff of 67.5 dBnHL revealed, respectively, a sensitivity of 75% and 70.6% and a specificity of 69.4% and 100%. For oVEMP (500 Hz), an amplitude cutoff of 10.8 μV or a threshold cutoff of 77.5 dBnHL revealed a sensitivity of 83.33% and a specificity of 87.5% and 80%, respectively. oVEMP (4 kHz), an amplitude cutoff of 3.1 μV, revealed a high specificity of 100% but a low sensitivity of 47.2%. A positive correlation was noted between the length of the SSCD and the cVEMP and oVEMP (500 Hz) thresholds and cVEMP amplitude. Conclusions: Our results support the use of oVEMP in the identification of SSCD. The presence of oVEMP (500 Hz) with an amplitude higher or equal to 10.8 μV, a threshold lower or equal to 77.5 dBnHL or oVEMP (4 kHz) amplitude of 3.1 μV represents the most useful to identify SSCD.