Sensors (Aug 2020)

Intrinsic Spectral Resolution Limitations of QEPAS Sensors for Fast and Broad Wavelength Tuning

  • Jesper B. Christensen,
  • Lasse Høgstedt,
  • Søren M. M. Friis,
  • Jui-Yu Lai,
  • Ming-Hsien Chou,
  • David Balslev-Harder,
  • Jan C. Petersen,
  • Mikael Lassen

DOI
https://doi.org/10.3390/s20174725
Journal volume & issue
Vol. 20, no. 17
p. 4725

Abstract

Read online

Quartz-enhanced photoacoustic sensing is a promising method for low-concentration trace-gas monitoring due to the resonant signal enhancement provided by a high-Q quartz tuning fork. However, quartz-enhanced photoacoustic spectroscopy (QEPAS) is associated with a relatively slow acoustic decay, which results in a reduced spectral resolution and signal-to-noise ratio as the wavelength tuning rate is increased. In this work, we investigate the influence of wavelength scan rate on the spectral resolution and signal-to-noise ratio of QEPAS sensors. We demonstrate the acquisition of photoacoustic spectra from 3.1 μm to 3.6 μm using a tunable mid-infrared optical parametric oscillator. The spectra are attained using wavelength scan rates differing by more than two orders of magnitude (from 0.3 nm s−1 to 96 nm s−1). With this variation in scan rate, the spectral resolution is found to change from 2.5 cm−1 to 9 cm−1. The investigated gas samples are methane (in nitrogen) and a gas mixture consisting of methane, water, and ethanol. For the gas mixture, the reduced spectral resolution at fast scan rates significantly complicates the quantification of constituent gas concentrations.

Keywords